Appendix B – Calibration, Traceability, and Recalibration
Ophir Pyroelectric Heads
Table B-3 Maximum Measurement Error due to Wavelength
Coating Type
Wavelength Error
Broadband
Metallic
190 - 350nm
±2%
±2%
400 - 800nm
±2%
±2%
1064nm 0
0
2 - 3µm
±2%
±2%
10.6µm ±5% ±15%
Recalibration from a Known Source of Laser Energy
To recalibrate from a known source of laser energy:
1. Click the right mouse button and select Calibrate from the pop-up menu.
OR
Open the Functions menu and select Calibrate. The Adjust Calibration
Factor dialog box appears.
Factor dialog box appears.
Figure B-7 Adjust Calibration Factor Dialog Box – Metallic Head
Figure B-8 Adjust Calibration Factor Dialog Box – Broadband Head
StarLab User Guide