Appendix B – Calibration, Traceability, and Recalibration  
Ophir Pyroelectric Heads 
Table B-3 Maximum Measurement Error due to Wavelength 
Coating Type 
Wavelength Error 
 Broadband 
Metallic 
190 - 350nm 
±2% 
±2% 
400 - 800nm 
±2% 
±2% 
1064nm 0 
2 - 3µm 
±2% 
±2% 
10.6µm ±5%  ±15% 
Recalibration from a Known Source of Laser Energy 
To recalibrate from a known source of laser energy: 
1.  Click the right mouse button and select Calibrate from the pop-up menu.  
OR 
Open the Functions menu and select Calibrate. The Adjust Calibration 
Factor
 dialog box appears. 
 
Figure B-7 Adjust Calibration Factor Dialog Box – Metallic Head 
 
Figure B-8 Adjust Calibration Factor Dialog Box – Broadband Head 