Chapter 4 – Measuring with the Thermopile Head 
Configuring Measurement Settings in Power Mode 
Selecting the Measurement Mode 
To select the measurement mode: 
1.  Select Power or Energy from the Mode drop down list in the 
Measurement Parameters Area. The Thermopile screen fields displayed 
depend on the mode selected. 
 
Figure 4-1 Mode Drop Down List 
Configuring Measurement Settings in Power Mode 
This section explains how to select the laser wavelength and configure the range 
and an average reading in power mode.  
Selecting the Laser Wavelength 
Thermopile heads have different absorption at different wavelengths. To 
compensate for these differences, each head has been calibrated by a laser at 
several wavelengths. By choosing the correct laser wavelength, the correction 
factor for that wavelength is automatically introduced.  
To select the laser wavelength: 
1.  Select the laser wavelength from the Laser drop down list in the 
Measurement Parameters Area. 
 
Figure 4-2 Laser Drop Down List 
LP1 Heads 
For the new LP1 family of heads, the absorption of the detector coating varies 
somewhat according to wavelength. The correction curve for the absorber is 
stored in the head EEROM. This correction curve ensures that the power reading 
is correct at all laser wavelengths.  