Chapter 4 – Measuring with the Thermopile Head 
Calibration Factors 
Calibration Factors 
StarLab allows you to adjust power calibration factors, configure the response 
factor and adjust energy calibration factors. 
 
Warning:
 
Adjusting the calibration factor makes a permanent change in the head. 
It is strongly recommended that before making any change to the factor, 
the original factory setting is recorded separately. This will make it 
easier to restore the value to its original setting later if needed
.
 
Adjusting Power Calibration Factors 
The absorption of the various Ophir thermal absorbers can vary from disc to disc. 
Therefore, all Ophir absorbers are individually calibrated against NIST traceable 
standards. Ophir heads are individually laser-calibrated at several wavelengths, 
against a NIST calibrated standard meter. For more information on Ophir head 
calibration and traceability, refer to Appendix B – Calibration, Traceability, and 
Recalibration
StarLab offers two types of calibration:  
Overall Calibration – Changes the calibration at all wavelengths at 
once. Use Overall Calibration if your head is equipped with the Ophir 
CAL resistor. For more information on the Ophir CAL resistor, refer to 
Appendix B – Calibration, Traceability, and Recalibration
• 
•  Laser Specific Calibration – Changes the calibration at one specific 
laser wavelength. Use Laser Specific Calibration, unless your head is 
equipped with the Ophir CAL resistor. 
To adjust the power calibration factor: 
1.  Click the right mouse button and select Calibrate from the pop-up menu.  
OR 
Open the Functions menu and select Calibrate. The Adjust Calibration 
Factors
 dialog box appears. 
 
Figure 4-7 Adjust Calibration Factors Dialog Box 
StarLab User Guide 
39