Chapter 5 – Measuring with the Photodiode Head 
Configuring Measurement Settings 
To remove laser wavelengths: 
1.  Select Remove from the Wavelength drop down list in the Measurement 
Parameters Area. The Remove Favorite Wavelength dialog box appears, 
displaying the wavelengths listed in the Laser Wavelength drop down 
list. 
 
Figure 5-4 Remove Favorite Wavelength Dialog Box 
Note: Remove is only available if more than 1 wavelength is listed in the 
Wavelength drop down list. 
2.  Select the wavelength you wish to remove. You can only remove one 
wavelength at a time. 
3.  Click Remove to remove the selected wavelength and close the dialog 
box. The removed wavelength no longer appears on the Wavelength drop 
down list. 
Filter Settings 
The photodiode head is equipped with an optional filter that allows it to measure 
up to 300mW or more without saturating the detector. The exact maximum 
power is reached when the reading reaches full scale, or when the output current 
from the head reaches 1.3mA, whichever comes first. You can work with or 
without the filter, depending on which powers you wish to measure. 
To configure the instrument to measure when the filter is inserted: 
1.  Select In from the Filter drop down list in the Measurement Parameters 
Area.  
 
Figure 5-5 Filter Drop Down List 
To configure the instrument to measure when the filter is removed: 
1.  Select Out from the Filter drop down list in the Measurement Parameters 
Area. 
Note: Make sure to physically insert or remove the filter, according to the 
filter setting selected. Failure to do so will result in erroneous 
readings. 
StarLab User Guide 
46