Chapter 6 – Measuring with the Pyroelectric and PD10 Heads 
Configuring Measurement Settings 
Using a Diffuser 
You can add a diffuser to some pyroelectric heads. A diffuser enables the energy 
ranges to reach very high levels. When using a head that can have a diffuser, 
configure the diffuser setting in the Measurement Parameters Area. 
To select whether a diffuser is in use:  
1.  Select whether a diffuser is in use (In) or not in use (Out) from the 
Diffuser drop down list in the Measurement Parameters Area. 
 
Figure 6-3 Diffuser Drop Down List 
When you change this setting, a dialog box appears reminding you to either 
insert or remove the diffuser. 
Note: For heads that are not equipped with a diffuser, this setting is 
unavailable, and the heads are configured to measure in Diffuser Out 
mode. 
Selecting the Pulse Width 
Some pyroelectric heads can be configured to measure long as well as short 
pulses. To accomplish this, the user must configure the head for long laser pulses 
or short pulses. 
Note: If the pulse length is incorrectly set to the short setting for pulses 
longer than that value, the reading will be erroneously low. If it is set 
to the longer setting for short pulses, the reading will be correct, but 
noisier. 
To select a pulse width: 
1.  Select the pulse width from the Pulse Width drop down list in the 
Measurement Parameters Area. 
 
Figure 6-4 Pulse Width Drop Down List 
Note: For heads with only one pulse width setting, the Pulse Width Drop 
Down list is unavailable and the head is configured to measure in its 
correct mode. 
StarLab User Guide 
55