Ophir Optronics Ltd QSR Manuel D’Utilisation
Chapter 6 – Measuring with the Pyroelectric and PD10 Heads
Adjusting Calibration Factors
Note: For metallic heads, when the calibration is changed at one laser
wavelength, the overall calibration of all other wavelengths changes
proportionately. For broadband heads, there is an option to adjust the
calibration factor for all wavelengths or only for a selected
wavelength.
proportionately. For broadband heads, there is an option to adjust the
calibration factor for all wavelengths or only for a selected
wavelength.
To adjust the energy calibration factor:
1. Click the right mouse button and select Calibrate from the pop-up menu.
OR
Open the Functions menu and select Calibrate. The Adjust Calibration
Factor dialog box appears.
Factor dialog box appears.
Figure 6-10 Adjust Calibration Factor Dialog Box – Metallic Head
Figure 6-11 Adjust Calibration Factor Dialog Box – Broadband Head
The Original area displays the original Measurement and Calibration
Factor. The Current wavelength is displayed beneath the Original area.
Factor. The Current wavelength is displayed beneath the Original area.
2. Use the Calibration Factor scroll bar to attain an accurate reading in the
Calibration Factor field in the Adjusted area.
OR
Enter the desired factor into the text box above the scroll bar.
3. Click Save to save the adjustment for the active channel.