Chapter 6 – Measuring with the Pyroelectric and PD10 Heads 
Adjusting Calibration Factors 
Note: For metallic heads, when the calibration is changed at one laser 
wavelength, the overall calibration of all other wavelengths changes 
proportionately. For broadband heads, there is an option to adjust the 
calibration factor for all wavelengths or only for a selected 
wavelength. 
To adjust the energy calibration factor: 
1.  Click the right mouse button and select Calibrate from the pop-up menu.  
OR 
Open the Functions menu and select Calibrate. The Adjust Calibration 
Factor
 dialog box appears. 
 
Figure 6-10 Adjust Calibration Factor Dialog Box – Metallic Head 
 
Figure 6-11 Adjust Calibration Factor Dialog Box – Broadband Head 
The Original area displays the original Measurement and Calibration 
Factor
. The Current wavelength is displayed beneath the Original area. 
2.  Use the Calibration Factor scroll bar to attain an accurate reading in the 
Calibration Factor field in the Adjusted area.  
OR 
Enter the desired factor into the text box above the scroll bar. 
3.  Click Save to save the adjustment for the active channel. 
StarLab User Guide 