VIVIX-S 1012N User Manual 
6.5 
Direct Calibration 
6.5.1  Prepration 
 
Precheck the state of X-ray generator and console. 
 
Precheck the state of X-ray tube. 
 
 
 
It is recommended you to check if X-ray dose value of the generator is exact by using the 
device like a dose meter. 
 
 
 
Make sure to preheat the detector for 30 minutes before starting calibration. The result of 
measurement can be incorrect if the detector is not preheated sufficiently. 
6.5.2  Offset Calibration 
 
 
1  Set the number of Stage as 5 in the Offset Calibration area. 
2  Click Run Offset Calibration button and progress the Offset calibration.