7.1.2 Battery Pack Diagnosis
Install the battery pack to the detector and check the voltage and remaining amount of
the battery pack. Furthermore, always check the remaining amount of the battery pack
during use of the detector. If performance of the battery pack has some problems, consult
the sales representative in Vieworks or a relevant engineer.
the battery pack. Furthermore, always check the remaining amount of the battery pack
during use of the detector. If performance of the battery pack has some problems, consult
the sales representative in Vieworks or a relevant engineer.
The battery pack belongs to consumables which performance will be decreased as time
passed. Make sure to check the battery life during usage.
passed. Make sure to check the battery life during usage.
Check remaining amout of battery pack
The battety remains is noticed as a level or percentage (%), and you can check the remains from VXvue
(Vieworks Viewer) or VIVIX SDK. It is also available to check the battery remains from the LED on the side
of detector.
(Vieworks Viewer) or VIVIX SDK. It is also available to check the battery remains from the LED on the side
of detector.
If the remaining of battery is under 30% or at the 1
st
level:
The releated warning message is displayed from VXvue.
The releated warning message is displayed from VIVIX SDK.
The detector LED for noticing battey remains is changed from green to orange color.
If the remaining of battery is under 10%:
The detector LED for noticing battey remains blinks in orange color.
You can check the battery status from the Diagnosis dialogue in VIVIX Setup program.
Display of Bettery Remains
Level
Value
Battery Remains
LED Display
Full
5
81% ~ 100%
5
th
level
Half Quarter
4
61% ~ 80%
4
th
level
Half
3
51% ~ 60%
3
rd
level
Quarter
2
31% ~ 50%
2
nd
level
Low
1
11% ~ 30%
1
st
level
1% ~ 10%
Blink
Unknown
0
Off