7.1.2  Battery Pack Diagnosis 
 
 
Install the battery pack to the detector and check the voltage and remaining amount of 
the battery pack. Furthermore, always check the remaining amount of the battery pack 
during use of the detector. If performance of the battery pack has some problems, consult 
the sales representative in Vieworks or a relevant engineer. 
 
 
 
The battery pack belongs to consumables which performance will be decreased as time 
passed. Make sure to check the battery life during usage. 
 
Check remaining amout of battery pack 
The battety remains is noticed as a level or percentage (%), and you can check the remains from VXvue 
(Vieworks Viewer) or VIVIX SDK. It is also available to check the battery remains from the LED on the side 
of detector. 
 
 
If the remaining of battery is under 30% or at the 1
st
 level: 
 
The releated warning message is displayed from VXvue
 
The releated warning message is displayed from VIVIX SDK
 
The detector LED for noticing battey remains is changed from green to orange color. 
 
If the remaining of battery is under 10%: 
 
The detector LED for noticing battey remains blinks in orange color. 
 
 
You can check the battery status from the Diagnosis dialogue in VIVIX Setup program. 
 
 
Display of Bettery Remains 
Level 
Value 
Battery Remains 
LED Display 
Full 
81% ~ 100% 
  5
th
 level 
Half Quarter 
61% ~ 80% 
  4
th
level 
Half 
51% ~ 60% 
  3
rd
 level 
Quarter 
31% ~ 50% 
  2
nd
 level 
Low 
11% ~ 30% 
  1
st
 level 
1% ~ 10% 
  Blink 
Unknown 
 
Off