Vieworks Co. Ltd. FXRD-1012NAW Manuel D’Utilisation

Page de 151
 
 
Rev.1.0
 
Page 132 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
7  Replace corresponding devices. 
8.1.5  Errors in Detector LED 
 
Symptom 
 
All LED lamps of detector are blinking. 
 
2 LED lamps of detector are blinking and remaining is blinking slowly. 
 
Expected Causes 
 
Internal hardware errors of the detector. 
 
Solutions 
1  Boot up the detector again and check the result of it 
2  Replace the detector. 
8.1.6  Rapid Consumption of Battery 
 
Symptom 
 
Consumption of fully charged battery pack is fast. 
 
Expected Causes 
 
Performance decrease caused by usage of long time. 
 
Usage of battery pack in low temperature environment. 
 
Solutions 
1  Replace to new battery pack if the battery pack has been used for a long time. (Battery pack is 
consumables) 
2  Use battery pack in room temperature environment. Charging capacity of battery pack in low 
temperature environment will be decreasing. 
8.1.7  Battery Pack or Installation Part of Battery is Getting Hot 
 
Symptom 
 
Battery pack or compartment for installation of battery pack is getting hot. 
 
Expected Causes 
 
Battery pack failure 
 
Detector Failure 
 
Solutions 
1  Do not use battery pack 
2  Consult with service engineers of Vieworks.