Vieworks Co. Ltd. FXRD-1012NAW Manuel D’Utilisation

Page de 151
 
 
Rev.1.0
 
Page 147 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
 
Power Frequency (50/60  ㎐) Magnetic Field IEC 61000-4-8 
Item 
Description 
Immunity test 
 
Power frequency (50/60  ㎐) magnetic field IEC 61000-4-8 
IEC 60601 test condition 
 
3 A/m 
Compliance Level 
 
3 A/m 
Electromagnetic Environment - 
Guidance 
 
Power frequency magnetic fields should be at levels characteristic of a 
typical location in a typical commercial or hospital environment. 
 
Conducted RF IEC 61000-4-6 / Radiated RF IEC 61000-4-3 
Item 
Description 
Immunity test 
 
Conducted RF IEC 61000-4-6      Radiated RF IEC 61000-4-3 
IEC 60601 test condition 
 
3 Vrms 150  ㎑  to 80  ㎒ 
    3 V/m 80  ㎒  to 2.5  ㎓ 
Compliance Level 
 
3 Vrms 150  ㎑  to 80  ㎒   
    3 V/m 80  ㎒  to 2.5  ㎓ 
Electromagnetic 
Environment - Guidance 
 
Portable and mobile RF communications equipment should be used no 
closer to any part of the EUT, including cables, than the recommended 
separation distance calculated from the below equations applicable to the 
frequency of the transmitter. 
     
   
 
 
P is the maximum output power rating of the transmitter in watts (W) 
according to the transmitter manufacturer and d is the recommended 
separation distance in meters (m). 
 
Field strengths from fixed RF transmitters, as determined by an 
electromagnetic site surveya, should be less than the compliance level in 
each frequency range b. 
 
Interference may occur in the vicinity of equipment marked with the symbol. 
 
 
 
 
At 80  ㎒  and 800  ㎒, the higher frequency range applies. 
 
These guidelines may not apply in all situations. Electromagnetic propagation is affected 
by absorption and reflection from structures, objects and people. 
 
 
 
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 
environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measured field strength in the location in which EUT is used exceeds 
the applicable RF compliance level above, EUT should be observed to verify normal 
operation. If abnormal performance is observed, additional measures may be necessary, 
such as reorienting or relocating EUT. 
 
Over the frequency range 150  ㎑  to 80  ㎒, field strengths should be less than [V1] V/m.