VIVIX-S 1012N User Manual 
1.3.5  Environment of Use 
 
 
Do not install the equipment in any of the locations listed below. Doing so may result in 
failure or malfuction, equipment falling, fire, or injury. 
 
Close to facilities where water is used. 
 
Where it will be exposed to direct sunlight 
 
Close to the air outlet of an air-conditioner or ventilation equipment 
 
Nearby the electric heating applicance such as a heater 
 
Where the power supply is unstable 
 
In a dusty environment 
 
In a saline or sulfurous environment 
 
Where temperature or humidity is higher than the operating temperature 
 
Where there is freezing or condensation 
 
In areas prone to vibration 
 
On an incline or in an unstable area 
 
This product may malfunction due to electromagnetic interference (EMI) caused by 
telecommunication devices, transceivers, electronic devices, etc. To prevent the 
electromagnetic wave from badly influencing the product, be sure to avoid placing it in 
close proximity to the product. Or, change direction or position of the product or move 
into the shielded place to reduce electromagnetic interference. 
 
This equipment is not suitable for use in the presence of a flammable anesthetic mixture 
with air or with oxygen or nitrous oxide. 
 
Conductive fluids that drain into the active circuit components of the system may cause 
short circuits that can result in electrical fire. Therefore, do not place fluids or food on any 
part of the system. 
 
To avoid electric shocks and burns caused by use of the wrong type of fire extinguisher, 
make sure that the fire extinguisher at the site has been approved for use on electrical 
fires. 
 
The battery charger and SCU (System Control Unit) cannot be used in patient’s vicinity.