Vieworks Co. Ltd. FXRD-1012NAW Manuel D’Utilisation

Page de 151
 
 
Rev.1.0
 
Page 26 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
2.3 
VIVIX-S 1012N Detector 
VIVIX-S 1012N is designed to acquire digital images by collecting x-ray signals and sereval conversion 
processes. You can use the acquired image (10” x 12” film size) diversely depending on the purpose of use. 
 
 
 
VIVIX-S 1012N is one of the VIVIX detector models produced by Vieworks. 
2.3.1  Specifications 
Item 
Specifications 
Model 
 
FXRD-1012NA (CsI) 
 
FXRD-1012NAW (CsI) 
 
FXRD-1012NB (Gadox) 
 
FXRD-1012NBW (Gadox) 
Image Sensor 
 
TFT: a-Si (Amorphous Silicon) 
X-ray Scintillator Type 
 
FXRD-1012NA(W) : Csl: TI (Thallium doped Caesuim Iodide) 
 
FXRD-1012NB(W) : Gd
2
O
2
S:Tb (Gadolinium oxysulfide) 
Pixel Pitch 
 
0.124㎜  (124㎛) 
Field of View 
 
10” x 12” 
Active Area (H x V) 
 
253.95㎜  ×  317.44㎜ 
Active Array 
 
2048 x 2560 pixels 
Effective Area 
 
252.5㎜  x 316.0㎜ 
Effective Array 
 
2036 x 2548 
Grayscale 
 
16bit 
Spatial Resolution 
 
Min. 4.0 lp/㎜ 
Image Acquisition Time (Wired) 
 
1.5 sec. 
Image Acquisition Time (Wireless) 
 
3 sec. 
Recommended Cycle Time 
 
15 sec. 
X-ray Synchronization Control 
 
AED (Auto Exposure Detection) 
 
DR Trigger (External line trigger) 
Rated Power Supply 
 
DC +24V, Max. 0.8A 
 
Wired: Powered by SCU with a tether interface cable. 
 
Wireless: Powered by a battery pack (3,100  ㎃ h) 
Power Consumption 
 
Max. 19.2 W 
Dimensions (H × W × D) 
 
287.0㎜  × 350.0㎜  × 15.0㎜ 
Weight (including a battery pack) 
 
2.2 kg 
Image Transfer 
 
Wired: Gigabit Ethernet (1000BASE-T) via PoE (Power over 
Ethernet) 
 
Wireless: IEEE802.11n 
Data Transmission Rate (Wired) 
 
Max. 1Gbps 
Data Transmission Rate (Wireless) 
 
Max. 300Mbps (MIMO 2X2)