2.3
VIVIX-S 1012N Detector
VIVIX-S 1012N is designed to acquire digital images by collecting x-ray signals and sereval conversion
processes. You can use the acquired image (10” x 12” film size) diversely depending on the purpose of use.
processes. You can use the acquired image (10” x 12” film size) diversely depending on the purpose of use.
VIVIX-S 1012N is one of the VIVIX detector models produced by Vieworks.
2.3.1 Specifications
Item
Specifications
Model
FXRD-1012NA (CsI)
FXRD-1012NAW (CsI)
FXRD-1012NB (Gadox)
FXRD-1012NBW (Gadox)
Image Sensor
TFT: a-Si (Amorphous Silicon)
X-ray Scintillator Type
FXRD-1012NA(W) : Csl: TI (Thallium doped Caesuim Iodide)
FXRD-1012NB(W) : Gd
2
O
2
S:Tb (Gadolinium oxysulfide)
Pixel Pitch
0.124㎜ (124㎛)
Field of View
10” x 12”
Active Area (H x V)
253.95㎜ × 317.44㎜
Active Array
2048 x 2560 pixels
Effective Area
252.5㎜ x 316.0㎜
Effective Array
2036 x 2548
Grayscale
16bit
Spatial Resolution
Min. 4.0 lp/㎜
Image Acquisition Time (Wired)
1.5 sec.
Image Acquisition Time (Wireless)
3 sec.
Recommended Cycle Time
15 sec.
X-ray Synchronization Control
AED (Auto Exposure Detection)
DR Trigger (External line trigger)
Rated Power Supply
DC +24V, Max. 0.8A
Wired: Powered by SCU with a tether interface cable.
Wireless: Powered by a battery pack (3,100 ㎃ h)
Power Consumption
Max. 19.2 W
Dimensions (H × W × D)
287.0㎜ × 350.0㎜ × 15.0㎜
Weight (including a battery pack)
2.2 kg
Image Transfer
Wired: Gigabit Ethernet (1000BASE-T) via PoE (Power over
Ethernet)
Ethernet)
Wireless: IEEE802.11n
Data Transmission Rate (Wired)
Max. 1Gbps
Data Transmission Rate (Wireless)
Max. 300Mbps (MIMO 2X2)