Vieworks Co. Ltd. FXRD-1012NAW Manuel D’Utilisation

Page de 151
 
 
Rev.1.0
 
Page 90 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
 
 
If Power Off is set to by SCU, the battery cannot supply power to the detector. In this 
case, you should connect the tether interface cable to supply power of SCU to the 
detector. 
 
Wireless network setting should be done by an engineer who understands the wireless 
communication and its related technique. Unless the network is set properly, a 
communication error would occur or the image quality would be affected. 
5.3.2  Detector Power Save Function 
The consumption of a battery pack can be reduced by using the power save function. 
 
 
 
The power save function is operated only when the battery supplies power to the 
detector. In other words, the power save function cannot be operated if a tether interface 
cable supplies power to the detector. 
 
Mode 
Mode 
Meaning 
Normal 
The detector can be operated and take an image at any time. 
Sleep 
The detector cannot be operated. User can take an image by disabling the Sleep mode. 
Shut Down 
The detector has been turned off. User can take an image after the detector is rebooted. 
 
 
Entry Condition of Power Save Mode 
Mode 
Meaning 
Normal 
Sleep 
The detector turns to sleep mode if not used for the setting time (Sleep after). 
Shut Down 
The detector is turned off if not used for the setting time (Shutdown after) under the 
sleep mode. However, if the detector is not used for the setting time (Shutdown after) 
under the sleep off state, the detector is turned off. 
 
 
Checks Power Save Mode 
Mode 
Meaning 
Normal 
All LED lamps are turned on. 
Sleep 
Power LED (Green) is blinking. 
VXvue (Vieworks Viewer) indicates the state of sleep mode. 
VIVIX SDK indicates the state of sleep mode. 
Shut Down 
All LED lamps are turned off.