Advanced Neuromodulation Systems Inc. dba. St. Jude Medical Neuromodulation Division SJMX599EPG Manuel D’Utilisation

Page de 33
 
30 
 
 
NOTE: 
Clean the EPG before attaching the header cap. 
ƒ
Clean the EPG by thoroughly wiping off its outer surface using the provided cleaning cloths. Do not 
wipe over the stimulator port. Do not submerge the EPG or use a cloth that is soaking wet. 
ƒ
Never use ammonia, acidic, or alkaline cleaners or organic chemicals such as paint thinner, 
acetone, propyl alcohol, or kerosene. They may damage surface finishes of the EPG. 
 
Storing and Transporting the EPG 
ƒ
Store the EPG and batteries in a cool, dry place. Do not store the EPG or batteries in your car or 
garage. 
ƒ
Transport the EPG in a carrying case. 
ƒ
If you are carrying the EPG in a briefcase or other carrying case, make sure no objects in the case 
are pressing on the device. 
ƒ
Store paired EPGs and patient controllers together to reduce the number of times you need to set 
up communication. 
ƒ
At the end of a trial, disable stimulation, clean the EPG, and attach the header cap before storing 
the EPG. 
37-5284-015.pdf   36
10/3/2014   3:40:06 PM