Monitor Mounting/Installation Options
M310 Fall Monitor – Setup & User Guide 17
Nurse Call Alarms
The Nurse Call jack can be used in all 3 Alarm volume settings (1, 2, 
3, and NC). When a Nurse Call jack is plugged in and the monitor’s 
alarm volume is set to “NC”, alarms are muted on the monitor. The 
“NC” volume setting is treated as a 1 setting if no Nurse Call jack is 
connected.
Note: 
An alarm sent to the Nurse Call system cannot be 
cancelled from the monitor alone. After dealing with 
the patient/resident according to your care policy, the 
alarm must first be cancelled on the monitor, and then 
at the Nurse Call system. Refer to the Nurse Call system 
documentation for details on cancelling alarms at the 
Nurse Call station.
Whether or not the volume setting is “NC”, the monitor sounds 
a beep when the Nurse Call jack is inserted or removed while the 
patient/resident is on the pad.
For more information on disabling the monitor, see “Hold Mode” on 
page 9.