Rehabtronics Inc. HSST1 Manuel D’Utilisation
20
Main Menu
When the STIM is turned on by pressing and holding the ON/OFF BUTTON, a Main Menu is presented
with the following four options:
with the following four options:
Menu Option
Description
STIM INTENSITY*
Adjust stimulation intensities to the electrodes controlling hand-open, hand-
close and thumb-close.
close and thumb-close.
FES MODE*
Functional Electrical Stimulation. Manually control hand-open and hand-close
stimulation.
stimulation.
EXERCISE MODE*
Timer-based stimulation programs that automatically cycle between hand-open
and hand-close stimulation.
and hand-close stimulation.
DEVICE SETUP
Settings to configure the user interface.
*NOTE: STIM INTENSITY, FES MODE and EXERCISE MODE stimulation functions require the STIM to be
connected to a GARMENT and armed for stimulation.
connected to a GARMENT and armed for stimulation.
STIM INTENSITY
Selecting the STIM INTENSITY option from the main-menu will access the following sub-menus:
Sub-Menu Description
Instructions
Resets all electrode stimulation
intensities to OFF.
intensities to OFF.
Press
twice to perform reset. Press any
other NAVIGATION BUTTON to cancel.
Adjust the stimulation intensity to the
electrode controlling hand opening, or
turn it OFF.
electrode controlling hand opening, or
turn it OFF.
Use the
and
buttons to adjust the
intensity.
Adjust the stimulation intensity to the
electrode controlling hand closing, or
turn it OFF.
electrode controlling hand closing, or
turn it OFF.
Use the
and
buttons to adjust the
intensity.
Adjust the stimulation intensity to the
electrode controlling thumb closing, or
turn it OFF.
electrode controlling thumb closing, or
turn it OFF.
Use the
and
buttons to adjust the
intensity.
NOTE: Increasing intensities requires the STIM to be connected to a GARMENT. The person wearing the
GARMENT will be able to feel the adjusted intensity.
GARMENT will be able to feel the adjusted intensity.
CAUTION: If sharing a single STIM between users, use the “reset” feature to reset all
electrode intensities to ‘OFF’ between users.
electrode intensities to ‘OFF’ between users.