Rehabtronics Inc. HSST1 Manuel D’Utilisation
23
DEVICE SETUP
Selecting the DEVICE SETUP option from the main-menu allows access to the following device settings:
DEVICE SETUP
Menu
Purpose
Instructions
Audio Setting
Enable or disable
audible beeps.
audible beeps.
Use the ‘<’ and ‘>’ navigation buttons to select the
preferred audio setting.
preferred audio setting.
Tap
Sensitivity
Adjust the tapping
force required to
cycle hand-
open/close/relax in
FES mode.
force required to
cycle hand-
open/close/relax in
FES mode.
Use the ‘<’ button to make the STIM less sensitive to taps
(harder tap required)
(harder tap required)
Use the ‘>’ button to make the STIM more sensitive to
taps (softer tap required).
taps (softer tap required).
The current setting can be tested by tapping the STIM.
The dot will flash green if the tap was hard enough for the
STIM to detect it at the current sensitivity setting.
The dot will flash green if the tap was hard enough for the
STIM to detect it at the current sensitivity setting.
HMS Mode
Select the preferred
head-motion
detection pattern
for use in FES mode.
head-motion
detection pattern
for use in FES mode.
Use the ‘<’ and ‘>’ navigation buttons to switch between
head-motion patterns.
head-motion patterns.
The currently selected pattern can be tested by donning a
connected HMS and performing the selected head-motion.
The dot will flash green when selected head-motion
patterns are detected.
connected HMS and performing the selected head-motion.
The dot will flash green when selected head-motion
patterns are detected.
Hand
Reorient the
DISPLAY and
NAVIGATION
BUTTONS according
to the affected arm.
DISPLAY and
NAVIGATION
BUTTONS according
to the affected arm.
Use the ‘<’ and ‘>’ buttons to select left or right hand
orientation, respectively. The DISPLAY and NAVIGATION
BUTTONS will re-orient automatically.
orientation, respectively. The DISPLAY and NAVIGATION
BUTTONS will re-orient automatically.
Usage Log
Displays statistics on
the amount of time
stimulating.
the amount of time
stimulating.
The Usage Log screen with automatically record and
update the following information:
update the following information:
o Hand Open – The total amount of time spent
stimulating the hand to open.
o Hand Close – The total amount of time spent
stimulating the hand to close.
o Total Cycles – The total number of times the hand has
been stimulated through open/close/relax cycles.