Wetting the ELECTRODES
The ELECTRODES must be wet with clean water prior to connecting them to the GARMENT. This is done
as follows:
as follows:
1. Inspect the ELECTRODES prior to use. If there is significant discoloration, bad odors, or any
noticeable separation of materials, discard and use a new set instead. For re-ordering
information “Contact and Reordering Information” at the end of this Manual.
information “Contact and Reordering Information” at the end of this Manual.
2. Soak ELECTRODES in clean, room-temperature water. Ensure the FABRIC PADS have
absorbed as much water as possible.
3. Squeeze ELECTRODES lightly to remove excess water. The ELECTRODES should be as wet as
possible, without dripping water.
CAUTION: Use only clean water to wet ELECTRODES. Using dirty water or any other liquid
can result in skin irritation and unknown stimulation performance.
can result in skin irritation and unknown stimulation performance.
Attaching the ELECTRODES to the GARMENT
Once the ELECTRODES have been wet, they can be snapped into the sockets at the ends of each of the
GARMENT’s ELECTRODE LEADS:
GARMENT’s ELECTRODE LEADS:
1. Snap the SMALL ELECTRODE into the socket of the GARMENT’s THUMB LEAD.
2. Snap the remaining three LARGE ELECTRODES in to the sockets of the remaining leads.
CAUTION: Ensure all four ELECTRODES are attached to the GARMENT prior to use. Wearing
the GARMENT with missing ELECTRODES can cause discomfort and skin irritation.
the GARMENT with missing ELECTRODES can cause discomfort and skin irritation.
CAUTION: Ensure each electrode is securely snapped into the GARMENT sockets. Loose
connections can result in intermittent stimulation delivery, or STIM error messages.
connections can result in intermittent stimulation delivery, or STIM error messages.