If irritation does not clear within an hour of removing the GARMENT, contact your healthcare 
professional. 
 
 
WARNING: If persistent skin irritation is experienced that does not clear within an hour of 
removing the GARMENT, discontinue use and contact your healthcare professional. 
 
 
Device Use Troubleshooting 
General Troubleshooting 
Issue 
Recommended Action(s) 
Device performance issues 
or unanticipated behavior. 
Discontinue use and contact customer support. 
(see “Contact and Reordering Information” at the end of this manual) 
 
STIM Error Messages 
The following table lists possible error messages that may appear on the STIM’s DISPLAY, and the 
recommended actions to take: 
DISPLAY 
Message 
Potential Cause(s) 
Recommended Action(s) 
A red X and a 
number 
- An internal hardware 
diagnostic test has failed 
Recharge the STIM and HMS in the DOCK
 
If the same error screen persists, take note of the 
number displayed and contact Rehabtronics support 
(see “Contact and Reordering Information” at the end 
of this manual). 
“Check 
Electrodes” 
Or 
“Impedance 
Error” 
ELECTRODES are not 
making good contact with 
the skin 
ELECTRODES are not 
sufficiently wet 
ELECTRODES are not 
properly snapped into the 
GARMENT 
Check that all ELECTRODES are sufficiently wet and 
pressed against the skin. 
 
Reattach the ELECTRODES to the GARMENT, ensuring 
that they are securely snapped to the sockets. 
“Put Stim in 
Garment” 
STIM is not in GARMENT 
STIM has been dislodged 
from the GARMENT 
Check that the STIM is firmly pushed into the 
GARMENT’s STIM HOLDER