Rehabtronics Inc. HSST1 Manuel D’Utilisation
44
If irritation does not clear within an hour of removing the GARMENT, contact your healthcare
professional.
professional.
WARNING: If persistent skin irritation is experienced that does not clear within an hour of
removing the GARMENT, discontinue use and contact your healthcare professional.
removing the GARMENT, discontinue use and contact your healthcare professional.
Device Use Troubleshooting
General Troubleshooting
Issue
Recommended Action(s)
Device performance issues
or unanticipated behavior.
or unanticipated behavior.
Discontinue use and contact customer support.
(see “Contact and Reordering Information” at the end of this manual)
(see “Contact and Reordering Information” at the end of this manual)
STIM Error Messages
The following table lists possible error messages that may appear on the STIM’s DISPLAY, and the
recommended actions to take:
recommended actions to take:
DISPLAY
Message
Potential Cause(s)
Recommended Action(s)
A red X and a
number
- An internal hardware
diagnostic test has failed
diagnostic test has failed
Recharge the STIM and HMS in the DOCK.
If the same error screen persists, take note of the
number displayed and contact Rehabtronics support
(see “Contact and Reordering Information” at the end
of this manual).
If the same error screen persists, take note of the
number displayed and contact Rehabtronics support
(see “Contact and Reordering Information” at the end
of this manual).
“Check
Electrodes”
Or
“Impedance
Error”
- ELECTRODES are not
making good contact with
the skin
- ELECTRODES are not
sufficiently wet
- ELECTRODES are not
properly snapped into the
GARMENT
making good contact with
the skin
- ELECTRODES are not
sufficiently wet
- ELECTRODES are not
properly snapped into the
GARMENT
Check that all ELECTRODES are sufficiently wet and
pressed against the skin.
Reattach the ELECTRODES to the GARMENT, ensuring
that they are securely snapped to the sockets.
pressed against the skin.
Reattach the ELECTRODES to the GARMENT, ensuring
that they are securely snapped to the sockets.
“Put Stim in
Garment”
- STIM is not in GARMENT
- STIM has been dislodged
from the GARMENT
- STIM has been dislodged
from the GARMENT
Check that the STIM is firmly pushed into the
GARMENT’s STIM HOLDER.
GARMENT’s STIM HOLDER.