Sensors & Software Inc. NG250 Manuel D’Utilisation

Page de 155
12-Survey & Map Mode
Noggin
94
The choices available are:
1)
Short
2)
Normal
3)
Long
4)
X-Long
5)
10x Normal
6)
20x Normal
7)
40x Normal
8)
50x Normal
9)
100x Normal
Each choice listed will be followed by an actual value in metres or inches depending on which
units are selected and which Noggin system is being used.  Here is a chart showing the station
interval for each system and setting.  Note the calculations for Data per Screen assumes that the
Plot Interval is set to Normal for the particular Noggin system (Noggin 250 = 2 pixels per trace
and Noggin 500 and 1000 = 1 pixel per screen).  If this assumption is not true, see the formula
after the charts for calculating this value.
N
OGGIN
 250 S
YSTEM
N
OGGIN
 500 S
YSTEM
Setting
Station Interval
Data per Screen
Short
  2.5 cm or   0.96 in
  8 m or 25.6 ft
Normal
  5.0 cm or   1.92 in
16 m or 51.2 ft
Long
10.0 cm or   3.84 in
32 m or 102.4 ft
X-Long
25.0 cm or   9.6 in
80 m or 256 ft
Norm x10
50.0 cm or 19.20 in
160 m or 512 ft
Norm x20
100.0 cm or 38.4 in
320 m or 1024 ft
Norm x40
200.0 cm or 76.8 in
640 m or 2048 ft
Norm x50
250.0 cm or 96.0 in
800 m or 2560 ft
Norm x100
500.0 cm or 192.0 in
1600 m or 5120 ft
Setting
Station Interval
Data per Screen
Short
  1.0 cm or 0.48 in
  6.4 m or 25.6 ft
Normal
  2.5 cm or 0.96 in
16 m or 51.2 ft
Long
  5.0 cm or 1.92 in
32 m or 102.4 ft
X-Long
12.5 cm or 4.8 in
80 m or 256 ft
Norm x10
25 cm or 9.6 in
160 m or 512 ft
Norm x20
50 cm or 19.2 in
320 m or 1024 ft
Norm x40
100 cm or 38.4 in
640 m or 2048 ft
Norm x50
125 cm or 48.0 in
800 m or 2560 ft
Norm x100
250 cm or 96.0 in
1600 m or 5120 ft