Care and Maintenance
Care and Maintenance
WARNING!
Perform maintenance tasks only when the device is 
unplugged.
Changing the
rechargeable
battery
Depleted rechargeable batteries can be replaced; 
contact BIOTRONIK.
Caution!
After a battery change, a complete battery 
maintenance cycle must be carried out so that the 
rechargeable battery reaches its full capacity and this 
can be displayed. Even if a full or partially charged 
battery is used, the maintenance cycle must be 
carried out (see the technical manual for the 
software).
Cleaning and Disinfection
To clean the device and the programming head, use a 
cloth moistened with a mild soap solution, alcohol, or 
other sterile solution (comply with the manufacturer’s 
recommendations concerning dilution before use).
Caution!
Never use organic solvents such as ether or acetone 
to clean the device. Always ensure that no liquids can 
penetrate the device.
Be careful not to allow any liquid to enter the housing 
during cleaning. 
Do not exert any pressure on the screen. Even slight 
pressure over a large area can irreparably damage the 
touch screen.
Disinfect the device with a mixture of 70% isopropanol 
and 30% water, or use Lysoformin 3000 (concentration 
= 2% , application time = 15 min).