PillCam Equipment Maintenance
PillCam Sensor Cleaning
205
PillCam Sensor Cleaning
This section includes instructions for cleaning the PillCam equipment. 
Cleaning the PillCam Sensor Belt
Cleaning the PillCam Sensor Array
For mild cleaning (dirt, sweat), wipe the sensors gently with alcohol wipes (up to 70%). The alcohol 
will not remove the adhesive. Use alcohol sparingly and allow the sensor array to dry for 20 minutes.
To remove adhesive from the sensor array (not from the human body), use white benzene.
Alternatively, use one of the following medical adhesive removers to remove adhesive:
B-508 Secure Solvent
B-202 Hollister Solvent
B-206 Detachol Adhesive Remover
Use all precautions as defined by the manufacturer.
Cleaning the Recorder Pouch
To clean the recorder pouch, wipe down all surfaces with alcohol (70% isopropyl or ethyl alcohol) 
making sure that all surfaces are exposed to alcohol for at least 1 minute.
֠
֠
֠
֠
Note
To clean the PillCam equipment surfaces, we recommend using alcohol 
wipes (up to 70%). Usage of other classes of disinfectants (Aldehydes, 
Oxidizing Agents, Quaternary Ammonium, Chlorine compounds, Iodophor, 
Phenolic compounds) may stain or damage the materials.
֠
֠
֠
֠
Note
Refer to the product insert supplied with your sensor belt for full instructions 
on fitting, usage, cleaning, and technical description.
!
Warning
Use white benzene only in a 
well ventilated
 area according to all 
precautions and instructions on the label.