Maintenance
Chapter 6
67
SensorBelt Cleaning
The SensorBelt may be machine washed after removal of the SensorBelt Insert
Follow instructions on the care label and use a mild detergent.
The surface of the SensorBelt may be wiped with any commonly used disinfectant.
The SensortBelt Insert may be wiped gently with alcohol (up to 70%). 
SensorArray Cleaning
For mild cleaning (dirt, sweat), wipe the sensors gently with alcohol (up to 70%). The alcohol 
will not remove the adhesive. Since alcohol is a polar solvent, do not use lavishly, and allow to 
dry for 20 minutes.
To remove adhesive from the SensorArray (not from the human body), use White Benzene. 
Note
White Benzene MUST be used in a ventilated area with all precautions 
defined in the manufacturer’s instructions.
Alternatively, use one of the following medical adhesive removers to remove adhesive:
B-508 Secure Solvent
B-202 Hollister Solvent
B-206 Detachol Adhesive Remover
Use all precautions as defined by the manufacturer.