How to Print Reports
HF PARTNER device over the patient’s implanted device to resume 
communication.
If the implanted device exhibited a lead or heart failure device fault, including 
when the implanted device’s battery is due for replacement, the fault detected 
light
 will stay on for five seconds after completion of the interrogation. 
Contact the physician who follows this patient’s implanted device. 
How to Print Reports
The HF PARTNER device prints a Diagnostic report that provides patient diagnostic 
data collected by the device. You may choose to print the last record or all records in 
the device memory.
Use the following steps to print desired reports:
1.
Place the HF PARTNER device near the printer adapter and press the appropri-
ate button sequence (Figure 4). To print the last record, press the blue PRINT 
button
once. To print all records press the blue PRINT button
twice. All 
lights will flash sequentially while the HF PARTNER device is looking for a printer. 
2.
When a printer is located, you will hear the “Connecting” message. The record 
transmission light
will blink while data is transmitting. Continue to keep the 
HF PARTNER device near the printer adapter. Upon successful data transfer you 
will hear the “Records transmitted” message and the report will print. 
3.
After printing is complete, the record transmission light
will stay on for 
five seconds.
Diagnostic Report
Last record requested
All records requested