Boston Scientific Corporation CRM312004 Manuel D’Utilisation
33
SPECIFICATIONS (Nominal)
Safety classification
Class I, Type CF, Defibrillation protected
Dimensions
19 in (48.3 cm) deep, 14.5 in (36.8 cm) wide,
4.75 in (12.1) cm high
4.75 in (12.1) cm high
Weight (approximate)
21.5 lb (9.8 kg)
Power rating
100–120/200–240 V and 50–60 Hz 3.8/1.9 A
Power cord
8 ft. 100–240 V. Reliable grounding is
achieved only when equipment is connected
to receptacle marked with “Hospital only” or
“Hospital grade.”
achieved only when equipment is connected
to receptacle marked with “Hospital only” or
“Hospital grade.”
Operating temperature
50–105
°F (10–40°C)
Transport and storage
temperature
temperature
–40 to 160
°F (–40 to 70°C)
Relative humidity
• Transport and storage
5–95%
• Operating
20–95%
Atmospheric pressure
500–1,060 hPa
External printer support
DB 25 parallel port connector
External monitor support
DB 15 VGA port connector
Analog output
+ 1 V output via seven-pin DIN connector
ECG performance
• Lead selection
I, II, III, aVR, aVL, aVF, V
• Intrinsic ventricular rate
display
30–240 bpm + 6 bpm
• Input impedance
> 2.5 M
Ω
• Electrode offset tolerance
300 mV
• Storage resolution
800 samples/sec, 4.56
µV
• Filter settings
ON: 0.5–25 Hz, + 0.2 dB;
OFF: 0.05–100 Hz + 0.2 dB
OFF: 0.05–100 Hz + 0.2 dB
• Gain settings
1, 2, 5, 10, 20 mm/mV + 25%
Internal printer
• Paper type
Thermo-sensitive
• Paper width
4 in (110 mm)
• Chart speed
10, 25, 50, 100 mm/sec
Battery type
DL 2450 or equivalent
Safety features
• Leakage current
< 10
µA (patient); < 100 µA (chassis)
• Defibrillator protection
to 5,000 V, 400 J