BEGIN ZIP TELEMETRY
NOTE:  The MICS telemetry feature is not available for all pulse generators. For more information, refer to the associated product literature for the pulse 
generator being interrogated.
1. 
Raise the PRM screen to a comfortable viewing angle.
2. 
Press the PRM On/Off  button.
3. 
Wait until both the PRM startup screen and the ZWT green indicator light appear, usually less than 60 seconds.
4. 
Place the telemetry wand over the pulse generator. Use either the Quick Start button on the PRM screen or the Select PG button located on 
the toolbar below the startup screen to begin interrogation. Remove the wand once the telemetry indicator  
  on the PRM is 
illuminated. 
For optimum MICS telemetry communication, position the ZWT at least 7.6 cm (3 in) away from the PRM. Remove any obstructions between the two 
instruments. Repositioning the ZWT may improve MICS telemetry performance. If MICS telemetry performance is not satisfactory, use the telemetry 
wand instead.
USING THE ZWT
Operation 
The ZWT requires special handling. To protect the ZWT from damage, refer to the following information:
• 
Do not subject the ZWT to abusive shocks or vibrations.
• 
When transporting the ZWT from an outside environment to an inside environment, allow the ZWT to come to ambient temperature before 
use.
• 
Do not place a magnet on the ZWT.
• 
Do not pour or splash liquid into or onto the ZWT.