Boston Scientific Corporation CRM314013 Manuel D’Utilisation

Page de 154
Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – 
guidance
Voltage dips, short interrup-
tions, and voltage variations 
on power-supply input lines 
(IEC 61000-4-11)
<5% 
U
T
 (>95% dip in 
U
T
 ) 
for 0.5 cycle
b
.  
40% 
U
T
 (60% dip in 
U
T
 ) for 
5 cycles
70% 
U
T
 (30% dip in 
U
T
 ) for 
25 cycles
<5%
U
T
 (>95% dip in 
U
T
 ) 
for 5 sec
<5% 
U
T
 (>95% dip in 
U
T
 ) 
for 0.5 cycle
b
.  
40% 
U
T
 (60% dip in 
U
T
 ) for 
5 cycles
70% 
U
T
 (30% dip in 
U
T
 ) for 
25 cycles
<5%
U
T
 (>95% dip in 
U
T
 ) 
for 5 sec
Mains power quality should be that of a typical 
commercial or hospital environment. If the user 
of the ZWT requires continued operation during 
power mains interruptions, it is recommended 
that the PRM to which the ZWT is connected be 
powered from an uninterruptible power supply 
or a battery.
Power frequency (50/60 
Hz) magnetic fi eld (IEC 
61000-4-8)
3 A/m
3 A/m
Power frequency magnetic fi elds should be at 
levels characteristic of a typical location in a typi-
cal commercial or hospital environment.
Note: 
U
T
 is the AC mains voltage prior to application of the test level.
16