NOTE:
  The handpiece is reusable and equipped with a disposable non-sterile 
 
             protective shield for single patient use.  The handpiece requires cleaning 
 
             before and after each patient treatment. For instructions on cleaning the  
 
             
the handpiece, refer to section 8.
 
 
NOTE:
  The Deep Tissue Handpiece is compatible only with the Re-useable Fiber 
 
             Optic Cable Assembly.  It is not compatible with the Classical (Feed-
 
               Through) Fiber.
 
 
 
 
 
Deep Tissue Handpiece 
 
 
1)  Remove Red Dust Cover from Deep Tissue Handpiece 
 
2)  Slide handpiece over monocoil shaft until it clicks
 
     into place.
 
 
3)  Place protective cover over the adjustable spacer 
 
4)  Loosen the Lock Ring and set the spacer at the desired spot size detent location.   
     Tighten the Lock Ring. 
 
5)  Place handpiece into the handpiece holder. 
 
 
 
 
To remove handpiece, press and hold the buttons on the side of the shaft and pull 
handpiece away from shaft. 
Classical “Feed-Through” Fiber Handpiece (Optional)