Biolase Inc. EPIC Manuel D’Utilisation
18 |
P a g e
5 4 0 0 3 2 1 R e v . X : D R A F T
NOTE:
The handpiece is reusable and equipped with a disposable non-sterile
protective shield for single patient use. The handpiece requires cleaning
before and after each patient treatment. For instructions on cleaning the
the handpiece, refer to section 8.
NOTE:
The Deep Tissue Handpiece is compatible only with the Re-useable Fiber
Optic Cable Assembly. It is not compatible with the Classical (Feed-
Through) Fiber.
Deep Tissue Handpiece
1) Remove Red Dust Cover from Deep Tissue Handpiece
2) Slide handpiece over monocoil shaft until it clicks
into place.
3) Place protective cover over the adjustable spacer
4) Loosen the Lock Ring and set the spacer at the desired spot size detent location.
Tighten the Lock Ring.
Tighten the Lock Ring.
5) Place handpiece into the handpiece holder.
To remove handpiece, press and hold the buttons on the side of the shaft and pull
handpiece away from shaft.
Classical “Feed-Through” Fiber Handpiece (Optional)