Biolase Inc. EPIC Manuel D’Utilisation
40 |
P a g e
5 4 0 0 3 2 1 R e v . X : D R A F T
Section 9: CALIBRATION
Calibration Schedule
Calibration procedure is recommended to be performed every 12 months in order to maintain
the required accuracy of output power versus displayed power. Annual calibrations can be
performed at a certified depot repair facility. Call Biolase Service at 1-800-321-6717 or your
Authorized Service Representative to schedule an appointment.
Calibration procedure is recommended to be performed every 12 months in order to maintain
the required accuracy of output power versus displayed power. Annual calibrations can be
performed at a certified depot repair facility. Call Biolase Service at 1-800-321-6717 or your
Authorized Service Representative to schedule an appointment.
Section 10: SOFTWARE SPECIFICATION
Biolase respects the intellectual property of others, and we ask our users to do the same. EPIC
10 software is protected by copyright and other intellectual property laws.
This product includes software developed by Biolase Technology Copyright ©2012 Biolase
Technology.
10 software is protected by copyright and other intellectual property laws.
This product includes software developed by Biolase Technology Copyright ©2012 Biolase
Technology.
Section 11: TROUBLESHOOTING
Below is the list of the Error messages, which in most cases can be fixed by the user.
If Corrective Action did not help, re-power the laser.
If Error is not cleared after re-powering, please call for system Service.
If Corrective Action did not help, re-power the laser.
If Error is not cleared after re-powering, please call for system Service.
NOTE: For all Error Messages not listed in the table, re-power the
system and if Error is not cleared, call for system Service.
Error/Warning Description
Reason
Fix
Thermistor Open
Thermistor Open
Thermistor Shorted
Thermistor Shorted
Call Biolase Service
Shutdown Temperature
System too hot
Allow 5-10 mins for laser to cool down
Laser Current High/ Low
Output is of specs
Call Biolase Service
FS shorted in Standby
FS not in Ready mode
Enter Ready mode
ON/OFF button Stuck
Key stuck
Press Front key
Flash Corrupted
Memory Corrupted
Call Biolase Service
No Fiber
Fiber not inserted
Plug in Trunk Fiber
Lost Footswitch Communication
No wireless connect
Re-establish pairing (see Sec 4)
Emergency Switch
E-Switch Pressed
Press E-Switch Again
Remote Interlock
Remote interlock open
Check Remote Interlock closed
Warning Temp High
System is hot
Allow 5-10 mins for laser to cool down
Warning Battery Low
Battery is low
Plug in DC supply
Warning Battery Not Connected
Battery not connected
Plug in Battery