Biolase Inc. EPIC-1 Manuel D’Utilisation

Page de 56
EPIC™10 User Manual 
pg. 22 
5400525 Rev B  
 
 
4.8 WIRELESS FOOTSWITCH 
The wireless footswitch is powered by two (2) AAA batteries.  
When the wireless footswitch is pressed in READY mode and the laser fires, a beeping sound 
indicates that laser energy is present.  A green LED will begin flashing and a blue LED will light 
at the top corners of the laser console, confirming the footswitch and laser are paired.  
In the top left corner of most screens is a Signal Strength Indicator 
 which displays the 
signal strength between the laser console and the footswitch (strongest is five (5) bars).  
Pressing and releasing the footswitch while in STANDBY mode will update this indicator.  
Although the unit will work with a signal level as low as one (1) bar, a weaker signal level will 
make the connection between the footswitch and laser console more vulnerable to wireless (RF) 
interference from other sources, such as cell phones or microwaves.  To improve the signal 
strength, reposition either the footswitch or the laser console until the signal indicator achieves 
the strongest possible level for optimal operation. 
NOTE: 
When the footswitch will go into SLEEP mode when not in use to conserve battery 
power. It automatically reactivates when it is pressed. 
 
4.9 PEAK POWER DISPLAY 
This number is shown only when the system is in Pulse mode and presents the value of the 
peak power based on the Power Setting and Pulse mode. 
4.10 PULSE MODE SELECTION 
Pulse mode selection graphically indicates whether the system is in Continuous mode or in 
Pulse mode. 
In Continuous mode, laser power is constantly delivered when the laser console is in READY 
mode and the wireless footswitch is activated. 
In Pulse mode, laser power is delivered in repetitive pulses, controlled by the Pulse Length and 
Pulse Interval settings.  Pressing the Pulse Mode button will allow switching between Pulsed 
and Continuous Modes (Figure 4.14). 
MODE
*
 
PULSE DURATION (on) 
PULSE INTERVAL (off) 
Duty Cycle 
(Time On / Time off) 
CP0 
10 microseconds 
40 microseconds 
20% 
CP1 
100 microseconds 
200 microseconds 
33% 
CP2 
1 millisecond 
1 millisecond 
50% 
P3 
20 milliseconds 
20 milliseconds 
50% 
    *CP = Comfort Pulse; P3 = Pulsed Mode which is the standard for most diode lasers currently available 
to the marketplace   
 
 
         Figure 4.13