Sun Nuclear Corporation 1136 Manuel D’Utilisation
Setup with control module
23
•
If the Pod number is unchanged or STOP is pressed, the display returns to the Main
screen.
screen.
PIN setup
This menu item is used to set up a PIN (Personal Identification Number). The administrator PIN
avoids accidental changes to setup and calibration. Default for the “Pin Setup” screen is 0000,
which is No PIN.
avoids accidental changes to setup and calibration. Default for the “Pin Setup” screen is 0000,
which is No PIN.
1
Press MENU button. The first menu item appears.
2
Continue pressing MENU button until “PIN Setup” screen appears.
3
Press ENTER.
4
Key in the 4-digit Administrator PIN number, and press
ENTER. If no PIN is set, enter 0000.
ENTER. If no PIN is set, enter 0000.
NOTE: If the PIN is lost or forgotten, the physicist can key in “911”
to display the stored PIN. After 911 is entered, fill in the stored PIN
and press ENTER to continue. Then the existing PIN can be used, or the PIN can be changed.
to display the stored PIN. After 911 is entered, fill in the stored PIN
and press ENTER to continue. Then the existing PIN can be used, or the PIN can be changed.
5
The “Key in new PIN #” screen appears.
6
Use the number keys to enter a 4-digit PIN. When entering
the numbers:
•
the numbers:
•
The arrow keys can be used to scroll left or right.
•
The un-initialized value is all 0’s and is invalid for a PIN.
•
All 4 digits must be used.
•
To remove a PIN, enter four zeros when prompted.
7
Press ENTER to confirm or STOP to abort without change.
The display returns to the Main screen.
The display returns to the Main screen.
Physics mode
Physics Mode is a special mode for the physicist to test and calibrate equipment. Administrator
pin is needed. Physics Mode allows the physicist to set up a variety of configurations for testing
and calibration.
pin is needed. Physics Mode allows the physicist to set up a variety of configurations for testing
and calibration.
To use Physics Mode:
1
Press MENU button. The first menu item appears.
2
Continue pressing MENU button until “Physics Mode”
screen appears.
screen appears.
3
Press ENTER.
4
Key in the 4-digit Administrator PIN number, and press ENTER.
5
Pressing the ENTER button turns ON Physics Mode and
momentarily displays the message “Physics mode is now
on.”
•
momentarily displays the message “Physics mode is now
on.”
•
The display returns to the Main screen in Physics Mode.
•
If the system is powered down while in the Physics Mode, it will automatically switch
back to normal mode when powered up again.
back to normal mode when powered up again.
•
If the Physics Mode is inactive for 1 hour, the system will automatically revert back to
standard mode.
standard mode.
NOTE: When Physics Mode is active, the letter 'P' appears in the
lower right hand corner of the dose display.
lower right hand corner of the dose display.
6
Pressing STOP turns the Physics mode OFF and momen-
tarily displays the message “Physics mode is now off.”
•
tarily displays the message “Physics mode is now off.”
•
To turn off Physics mode, press MENU, scroll to Physics Mode, enter PIN, and press
STOP; or, power down the system.
STOP; or, power down the system.
K e y i n p i n # _ _ _ _
t h e n E N T E R
K e y i n n e w P I N #
9 9 9 9
T h e n E N T E R
E N T E R = C o n f i r m
S T O P = A b o r t
P h y s i c s M o d e
P r e s s E N T E R
E N T E R = M o d e O N
S T O P = M o d e O F F
A = 9 . 9
B = 9 . 9
C = 9 . 9 c G y P