Setup for PC software
47
Figure 4-5. Temperature calibration view
Save Cal button—saves the entered/completed temperature calibration to the Pod.
Cancel Temperature Cal button—cancels the current temperature calibration (if the input is 
selected (Red Check mark).
Inputs to Temperature Calibrate—lists the diode serial numbers and previous temperature 
coefficients. To select an input for calibration, right-click the input. Once selected, a check 
mark appears next to the selected input. The diode serial numbers are carried over from the 
dose calibration, which has to be completed before temperature calibration can be done. 
Temperature factors can be entered directly from a chart for SNC detectors and certain other 
manufacturers. If this method is used, the temperature calibration does not have to be done 
which greatly simplifies this process. See “Temperature coefficient (Tc)” on page 62.
Checklist—lists the steps and the sequence of operation for a temperature calibration. This 
is your list of steps to perform a single calibration. During calibration, you click on each task 
when it is completed. A check mark appears in the box to identify the completed items. You 
must complete the checklist in the correct order from top to bottom.
Additional Instructions—lists additional instructions for the step to be completed.
Setup for PC software
COM (serial) connection
Connect the serial cable between the PC’s serial port and the COM port on the Control Mod-
ule (Figure 4-6).
NOTE: You can permanently install the 25-meter data cable in a conduit between the two rooms. 
For multiple rooms, additional 25-meter data cables can be purchased separately and installed in 
all conduits and control rooms. This will reduce setup time when moving the system from room to 
room. 
Connect the instrument to the Control Module with the power/data cable.
Plug the power converter into an outlet with 110 to 240 VAC, single phase, 50-60 Hz power. 
Connect the power converter to the Control Module.