Sun Nuclear Corporation 1136 Manuel D’Utilisation

Page de 98
Specifications
5
Figure 1-5. Using the BNC to LEMO adaptor
Specifications
Table 1-2. Specifications for IVD solutions 
Category
Characteristic
Specification
Diode detector inputs Quantity
Four channels; connection to 4 diodes
Connector type
LEMO (BNC-LEMO adaptor available)
Detector temperature 
Compensation
Auto, within 0.1 °C (negative detector current only)
Calibration
User calibrated against ion chamber standard
Repeatability
± 0.2% or ± 0.1 cGy
Detector requirements n or p type diodes responding between 5 to 100 nC/Gy
Polarity
Bipolar, i.e., measures positive or negative detector current
Current limit
± 40 nA average radiation current or 1.0 nC per radiation 
pulse
Leakage
Auto compensation
Warm-up time 
< 30 seconds
Environmental 
Temperature
15 °C to 32 °C 
Relative humidity
10% to 90% rh, non-condensing
Component defini-
tions, size and weight
Detector Pod
Measure dose received by diode detectors, 4 diodes per 
Pod; 7x12x3 cm, 340 gm
Base station
Data relay between wireless Detector Pod and Control Mod-
ule; 7x12x3 cm, 340 gm
Display
Control Module with keypad for start and display of dosim-
etry measurements; 14x16x3.5 cm, 780 gm
Printer
Seiko thermal printer, model DPU-414, serial port to display, 
9x320 dots/line, 40 column=9 cm; 16x17x6.6 cm, 580 gm 
Cable
Control Module to base station (located in treatment room) 
via 6 pin modular connector/cable, 25 m.
Frequency 
916.5 MHz for USA, 868.35 MHz for European Union
Power
Detector Pod
Rechargeable NiMH battery, 12 hour on full charge
Control Module
Medically-approved power supply operating on 100-240 
VAC, 47-63 Hz mains, providing 18 VDC output
Electrical isolation 
• Wireless—through radio frequency communication, Am-
plifier Sequenced Hybrid (ASH), On-Off Keyed (OOK), 
Manchester encoded
• Hardwired—through medically approved power supply