Warnings, Precautions and Notes
x
0070-10-0690-01
Panorama™ Operating Instructions
Warnings
WARNING: Only qualified and trained personnel or Datascope Service 
personnel should attempt to service Datascope equipment. 
Service is defined as any activity requiring the cover to be 
removed for internal adjustments, parts replacements, 
repairs or software upgrades of any kind to insure 
compatibility.
WARNING: To insure compatibility with the operating system and 
applications software, use only Datascope Corp. supplied 
and/or approved components to repair any part of the 
Panorama Patient Monitoring Network. Use of 
unauthorized software, devices, accessories, or cables other 
than those approved by Datascope may render the 
application unsuitable for medical patient monitoring. It 
may also result in increased electromagnetic emissions or 
decreased Immunity of the system.
WARNING: Route cables neatly. Ensure cables are not in the way of 
patient or hospital personnel.
WARNING: Do not incinerate batteries, possible explosion may occur.
WARNING: Some of the equipment on the Panorama Patient Monitoring 
Network must utilize the hospital emergency power system. 
Failure to do so will result in loss of monitoring during 
extended periods of power failure. The back-up power time 
period, for the Panorama Patient Monitoring Network, is 
limited.
WARNING: The 18.1" flat panel may tip over if the display head is 
inclined to an angle greater than 45° backward tilt. If the 
user elects to have the display head inclined to an angle 
greater than 45° backward tilt, the flat panel must be 
attached to a secure mounting surface via three screw 
locations on the bottom of the base.
WARNING: Do not put MPSO (Multiple Portable Socket Outlets, i.e., 
multiple outlet extension cords) used with the Panorama 
Patient Monitoring Network equipment on the floor. Connect 
only approved accessories and components to the same 
MPSO as the Panorama Patient Monitoring Network 
equipment. Do not overload MPSOs. Use only MPSOs that 
comply with the requirements of IEC 60601-1-1. 
WARNING: Incorrect setting or silencing of patient alarms can 
jeopardize patient safety.
WARNING: When the Suspend Non-Lethal alarms sidebar button is 
enabled, the Panorama Central Station will not monitor any 
non-lethal arrhythmias, and will not display numeric 
physiological or technical alarms for the selected patient tile. 
WARNING: When the Suspend All Alarms sidebar button is enabled, the 
Panorama Central Station will not monitor any lethal or 
non-lethal arrhythmias and will not display numeric 
physiological or technical alarms for the selected patient tile.