Mindray DS USA Inc. WT0707-00000 Manuel D’Utilisation
Panorama™ Operating Instructions
0070-10-0690-01
1 - 17
System Overview
Troubleshooting
Disabled Alarm Tone,
QRS Tone or Other
Function
QRS Tone or Other
Function
MUTE button pressed.
Check for alarm mute symbol and
message.
message.
Beep volume low.
Increase beep volume.
ECG Base Line With No
Wave form
Wave form
The Gain control is not set high
enough. Set through SIZE button.
enough. Set through SIZE button.
Readjust as required.
Lead wires and patient cable not
fully inserted into proper
receptacle.
fully inserted into proper
receptacle.
Check insertion.
Cable or lead wires damaged.
Check with lead continuity tester.
Base Line Wander
Patient moving excessively.
Secure lead wires and cable to
patient.
patient.
Patient's respiration.
Reposition electrodes
Electrodes dry.
Reprep skin and apply fresh moist
electrodes.
electrodes.
Static build up around patient.
Check with hospital engineer.
Damped Invasive Wave
form
form
Air bubbles in tubing.
Eliminate air from tubing.
Kinked catheter.
Slightly alter position of catheter.
Catheter against wall of blood
vessel.
vessel.
Check for leaks at connector.
Blood in tubing.
Pump pressure bag up to 300
mmHg.
mmHg.
No waveform for an
invasive BP
invasive BP
Improper attachment of
transducer to monitor.
transducer to monitor.
Check transducer connection.
Faulty transducer.
Try a different transducer.
Recorder Report is Blank
Thermal paper may be installed
incorrectly. (upside down)
incorrectly. (upside down)
Remove paper and reinstall with
paper feeding off of the spool
from the bottom.
paper feeding off of the spool
from the bottom.
Resp. Wave form Too
Large
Large
Scales set inappropriately.
Change lead selection.
Resp. Wave form Too
Small
Small
Patient breathing shallow or
turned on side.
Scale set inappropriately.
turned on side.
Scale set inappropriately.
Change lead selection
False Apnea Alarm
Apnea delay may be improperly
set. Patient may be having
frequent episodes of CVA.
set. Patient may be having
frequent episodes of CVA.
Choose another apnea delay.
Scale size may be too low.
Reposition electrodes to better
detect respirations.
detect respirations.
CHK Lead
Due to increased impedance.
Chest hair under electrodes.
Chest hair under electrodes.
Prep chest.
Dried electrode gel. Change
electrodes.
Electrode off.
Replace electrode.
Lead off.
Replace lead.
Cracked lead wires.
Replace lead wires.
Poor skin prep.
Clean and abrade skin before
applying electrodes.
applying electrodes.
CVA
Can be caused by shallow
breathing or an apnea event.
breathing or an apnea event.
Check the patient
Adjust scales or leads if necessary.
Adjust scales or leads if necessary.
MESSAGE/ISSUES *
REASON
SOLUTION
*
Messages are shown in all bold text.