Operator’s Manual  ZM-520PA/521PA/530PA/531PA
 51
Monitoring
CAUTION
The measurement values and displayed waveforms on the transmitter and receiving 
monitor may be different due to timing delay of the display or difference in detection 
settings.
NOTE
Do not let the transmitter continuously contact the patient’s skin directly. The transmitter 
heats up by 2 or 3°C (4 or 5°F) during operation and it may cause low temperature burn 
to the patient.
Screen Descriptions
When the transmitter is turned on, the startup screen appears, then the check electrodes screen to 
check the electrode attachment appears.
The screen changes in the following order each time the Screen key is pressed.
Check electrodes → numeric and waveform → waveform review → numeric review → display 
off → check electrodes . . . 
Screen key