Understanding Test Results and Messages
If your blood glucose result is unusually high or low, or if you question your test results, repeat the 
test with a new 
 Blood Glucose Test Strip GS700.
You can also run a Quality Control Test with the 
 Control Solutions GC700  to check your 
 Blood Glucose Meter
 and 
Blood Glucose Test Strip GS700 (Refer to 
“Performing a Quality Control Test” on page 40) .
If the test result still remains unusually high or low, contact your healthcare professional 
immediately.
If you are experiencing symptoms that are not consistent with your blood glucose test results and 
you have made sure to follow all instructions in this manual, contact your healthcare professional 
immediately.
Follow local regulations or consult your health care provider for appropriate disposal of used test
strips and lancets.
TM
 Rightest
TM
 Rightest
 
TM
TM
Rightest
Rightest
 
 
GM782
 
Make sure your blood sample covers the 
whole area of the view window to get an 
accurate test result. An insufficient blood 
sample will result in an error message 
("Er4"). If this occurs, repeat the test with a 
new test strip.
 Insufficient blood sample
 Sufficient blood sample
CAUTION
- Check the expiration date printed on the strip vial every time you use a test strip. Do not use 
expired 
 Blood Glucose Test Strips GS700.
- Use each 
 Blood Glucose Test Strip 
 immediately after removing it from 
the vial.
- Do not reuse 
 Blood Glucose Test Strips 
.
- Apply the blood sample only on the sample port of the test strip.
TM
Rightest
TM
Rightest
GS700
TM
Rightest
GS700
Do not perform testing immediately after moving from one ambient temperature to another 
(e.g. after coming indoors from the outside). Allow 30 minutes for the temperature of the 
meter and the test strips to equilibrate.