Guidance and manufacturer’s declaration – electromagnetic immunity 
The SF-371 body fat scale    is intended for use in the electromagnetic environment specified below. The customer 
or the user of the SF-371 body fat scale    should assure that it is used in such an environment. 
Immunity test 
IEC 60601 
test level 
Compliance level 
Electromagnetic environment - 
guidance 
Electrostatic 
discharge (ESD) 
 
IEC 61000-4-2 
± 6 kV contact 
 
± 8 kV air 
 
 
± 6 kV contact 
 
± 8 kV air 
 
 
Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic 
material, 
the 
relative 
humidity 
should be at least 30 %. 
Electrostatic 
transient / burst 
 
IEC 61000-4-4 
± 2 kV for power 
supply lines 
 
± 1 kV for input/output 
lines 
 
 
 
N/A 
 
 
 
Mains power quality should be that of 
typical commercial or hospital 
environment. 
Surge 
 
IEC 61000-4-5 
 
±    1 kV differential 
mode 
 
± 2 kV common mode 
 
N/A 
 
Mains power quality should be that of 
typical commercial or hospital 
environment. 
Voltage 
dips, 
short 
interruptions and 
voltage variations 
on power supply 
input lines 
 
IEC 61000-4-11 
< 5 % U
T
   
(>95 % dip in U
T
 ) 
for 0.5 cycle 
 
40 % U
 
(60 % dip in U
T
 ) 
for 5 cycles 
 
  70 % U
 
(30 % dip in U
T
 ) 
for 25 cycles 
 
< 5 % U
T
   
(>95 % dip in U
T
 ) 
for 5 sec   
N/A 
 
Mains power quality should be that of 
typical 
commercial 
or 
hospital 
environment. If the user of the   
SF-371  body  fat  scale      requires 
continued  operation  during  power 
mains 
interruptions, 
it 
is 
recommended  that  the  SF-371  body 
fat  scale      be  powered  from  an 
uninterruptible  power  supply  or  a 
battery. 
Power frequency 
(50/60 Hz) 
magnetic field 
 
3 A/m 
 
 
3 A/m 
 
Power  frequency  magnetic  fields 
should  be  at  levels  characteristic  of  a 
typical 
location 
in 
typical 
commercial or hospital environment.