TECNICAS CIENTIFICAS PARA LABORATORIO S.A. T620 Manuel D’Utilisation
ISEsweat
- User’s Manual 22
4.2 Sweat induction and analysis
PREVIOUS PHASE
Follow the instructions on the screen:
Insert sensor
1. Select a card and take it out of the package
2. Remove the protective film plastic that protect the
connectors. Dry any excess liquid from the
connections zone with a tissue
connections zone with a tissue
3. Before starting the analysis, locate the IONTO side
of the sensor card.
NOTE: The IONTO side of the card is designed for skin
stimulation, in order to obtain a sweat sample prior to
running the sweat test in the Normal Mode.
4. Place the sensor card with the IONTO side facing
stimulation, in order to obtain a sweat sample prior to
running the sweat test in the Normal Mode.
4. Place the sensor card with the IONTO side facing
outwards in the remote connector slot, as shown in
the image.
the image.
5. Remove the protective plastic protector strip on the
IONTO side.
CAUTION: Please, do not touch the hydrogels on the IONTO side; it could affect their
composition. Pilocarpine found in the hydrogels can be toxic if inhaled or ingested. Please,
clean carefully all contact zones after the analysis. In some cases, depending on the patients
sensitivity, the gels can cause an allergic reaction such as red spots on the skin which may
need to be treated with anti-inflammatory treatment. If the patient suffers any distress or pain,
STOP the test immediately and cancel the analysis
composition. Pilocarpine found in the hydrogels can be toxic if inhaled or ingested. Please,
clean carefully all contact zones after the analysis. In some cases, depending on the patients
sensitivity, the gels can cause an allergic reaction such as red spots on the skin which may
need to be treated with anti-inflammatory treatment. If the patient suffers any distress or pain,
STOP the test immediately and cancel the analysis
CAUTION: In case you introduce the card the wrong way round (opposite side of the one
needed); the remote LED will remain red. Extract the card, turn it round and introduce it again,
correctly. The LED will go out and you can continue the procedure.
needed); the remote LED will remain red. Extract the card, turn it round and introduce it again,
correctly. The LED will go out and you can continue the procedure.