-  User’s  Manual                                                                                 22 
 
 
 
 
4.2  Sweat induction and analysis 
 
PREVIOUS PHASE 
 
Follow the instructions on the screen:  
 
Insert sensor 
 
 
1.  Select a card and take it out of the package   
 
 
 
2.  Remove the protective film plastic that protect  the 
connectors. Dry any excess liquid from the 
connections zone with a tissue 
 
 
 
3.  Before  starting  the  analysis,  locate  the  IONTO  side 
of the sensor card.  
 
 
NOTE: The IONTO side of the card is designed for skin 
stimulation,  in  order  to  obtain  a  sweat  sample  prior  to 
running the sweat test in the Normal Mode.  
 
4.  Place  the  sensor  card  with  the  IONTO  side  facing 
outwards  in  the  remote  connector  slot,  as  shown  in 
the image. 
 
 
5.  Remove  the  protective  plastic  protector  strip  on  the 
IONTO side. 
 
 
 
 
 
 
 
CAUTION:  Please,  do  not  touch  the  hydrogels  on  the  IONTO  side;  it  could  affect  their 
composition.  Pilocarpine  found  in  the  hydrogels  can  be  toxic  if  inhaled  or  ingested.  Please, 
clean carefully all contact zones after the analysis. In some cases, depending on the patients 
sensitivity,  the  gels  can  cause  an  allergic  reaction  such  as  red  spots  on  the  skin  which  may 
need to be treated with anti-inflammatory treatment. If the patient suffers any distress or pain, 
STOP the test immediately and cancel the analysis 
 
CAUTION:  In  case  you  introduce  the  card  the  wrong  way  round  (opposite  side  of  the  one 
needed); the remote LED will remain red. Extract the card, turn it round and introduce it again, 
correctly. The LED will go out and you can continue the procedure.