2.
Discard the Patch immediately. Do not reuse.
If required, replace the Patch with a new one according to the instructions in Sections 6.2 to 6.4.
The Patch should be replaced at least once every five days. Make sure to place the new Patch
on the same spot (or as closely as possible to the location) of the previous Patch).
5.7
RECHARGING THE SENSOR
1.
The Sensor should be charged every three days, or
with every replacement of the Patch, the sooner of
the two.
2.
Place the Sensor in the Charger. See Figure 17.
The color of the light indicator on the Sensor will
be amber when being charged. It will turn green
when the Sensor is fully charged and ready to go.
5.8
SENSOR AND LIGHT INDICATOR STATUSES
The following is a summary of the light indicator statuses of the Sensor:
Sensor Position
Light Color
Meaning
Sensor in Charger
Solid green
Battery is fully charged and is ready for use.
The solid green light will continue until the
Sensor is removed from the Charger.
Solid amber
Battery is not full and is being charged. The
light will change to green once battery is
charged and Sensor is ready for use.
Sensor on the body
Solid green
Sensor is ready for monitoring. This solid
green light will last for several seconds, after
which it will disappear.
Blinking
amber
Sensor error. Call ZOLL customer service at
1.888.592.3798.
Figure 107: Sensor Charging