Innokas Yhtyma Oy VC150 Manuel D’Utilisation
KO00065K
VC150 Vital Signs Monitor
D-13
Supplemental analysis of clinical accuracy test data: Masimo sensor accuracy
Performance specifications for Rainbow ReSposable Pulse CO-Oximeter
Sensor System
Sensor System
Rainbow Adhesive and Resposable Sensors
The table below shows Arms (Accuracy Root Mean Square) values measured
using the Rainbow Adhesive and Resposable Sensors sensors with Masimo SET
Oximetry Technology in a clinical study.
using the Rainbow Adhesive and Resposable Sensors sensors with Masimo SET
Oximetry Technology in a clinical study.
Rainbow Adhesive and Resposable Sensors, measured values
Range
Arms
90-100%
1.25%
80-90%
1.45%
70-80%
1.94%
Error (SpO
2
- SaO
2
)
SaO
2