For setting “Type of Device” in the “Settings” panel (Menu—“Edit”--”Settings”), 
select “R-100(PTL) / R-300(PTL)” ,and other setting should be as follows.
 
Please note that these settings should be common with LTS-200's. And if the 
selection of “Type of Device” is not correct it may result in missing some data. 
LTS-200  →   PC(DL-01) 
Bits per second: 
1200 (1200-9600 for sending “POLAR DATA”) 
Databits: 
Parity: 
None 
Stop bits: 
PC(DL-01)  →  LTS-200 
Bits per second: 
1200 
Databits: 
Parity: 
None 
Stop bits: 
c) Note for the Memory capacity. 
Data transfer failure from DL-01 to LTS-200 may cause reduction of memory capacity. 
If memory capacity becomes less, please back up required data first, and then initialize 
coordinates data. 
To initialize coordinates data, turn on the instrument while pressing [F2]+[F5]+[ON/OFF], 
and take your finger off [ON/OFF] again. 
After you see the message “COORD. DATA INITIAL”, press [F5]. Then the message 
Please wait” is displayed. When it is completed, the panel of MODE A is displayed. 