testo Instruments Co. Ltd 05605522 Manuel D’Utilisation
5 Maintenance
22
5.2
Cleaning the instrument
Contaminants such as oil may impair the accuracy of the vacuum
sensor. Complete the following steps to clean the sensor.
sensor. Complete the following steps to clean the sensor.
CAUTION
Carrying out cleaning with the instrument switched on may result in
damage to the sensor!
damage to the sensor!
- Before cleaning, switch the instrument off!
CAUTION
Damage to the sensor due to sharp objects!
- Do not insert any sharp objects into the connections!
1
- Switch the instrument off.
2
- Put a few drops of rubbing alcohol into one of the two connections.
3
- Seal the opening by placing your finger on it or screw on the sealing
caps.
- Shake the instrument briefly.
4
- Remove all the alcohol from the instrument.
5
- Repeat this process at least twice.
6
- Leave the instrument to dry for at least 1 hour. To dry the sensor
faster, you can connect the probe directly to a vacuum pump and
draw vacuum.