Valeport Ltd 0804IDU Manuel D’Utilisation

Page de 83
 
HydrINS 2 / HydrINS 2 Mini flow meter  
Installation Manual  
 
Page 32 / 83 
 
 
 
 
Configuration with pulse 
output in bare wire to a 
SCADA system 
Configuration with a 
standalone recorder 
Configuration with 
Display (standalone or 
supplied from outside) 
Mode 1 
Mode 1 
Mode 2 
Figure 20: Installation configurations for HydrINS 2 / HydrINS 2 Mini 
It  is  essential  to  ensure  that  the  HydrINS  2  /  HydrINS  2  Mini  flow  meter  is  configured  in  the 
appropriate  operating  mode  before  placing  it  in  operation;  otherwise,  the  probe  will  suffer 
irreversible damage (see paragraph 8.2)
8  Installation of a display 
It  is  necessary  to  check  in  advance  that  the  HydrINS  2/HydrINS  2  Mini  flow  meter  is  configured  in 
Mode  2  (see  paragraph  8.23).  It  is  important  to  specify  that  there  is  always  the  possibility  of 
communicating  with  the  flow  meter  from  a  computer  to  perform  checks  or  acquisition 
reprogramming  because  the  display  is  'transparent'  in  communication  between  the  computer  and 
the HydrINS 2 / HydrINS 2 Mini flow meter. 
8.1  Preview of Displays 
The Displays have cable gland fittings and/or military connectors and/or a TNC connector. 
To reduce energy consumption, the display is activated manually using a magnet included with the 
Display. This magnet must be applied to the area represented by the symbol 
 on one of the sides of the housing (Figure 22), to activate the Display. The display time is 
adjustable when programming the probe. 
 NOTE: This point is essential before putting the unit in operation. If an 
incorrect operating mode is selected, the probe electronic board could 
suffer irreversible electrical damage.