Valeport Ltd 0804IDU Manuel D’Utilisation

Page de 83
 
HydrINS 2 / HydrINS 2 Mini flow meter  
Installation Manual  
 
Page 42 / 83 
8.5.3  Internal power supply using alkaline batteries 
1  –  Remove  all  connections  from  Connector  J15,  and  then  connect  the  additional  board,  Cat.  No., 
4005511  to  the  same  connector.  In  this case,  the  position  of  Switch  SW2  used  earlier  is  no  longer 
relevant. Switches SW1 and SW2 of the additional board will be used. 
2 - Create the following configuration:  
 
Switch SW1 of the additional board: Any position 
 
Switch SW2 of the additional board: Internal 
3 – Connect the battery pack to the +/- 9 to 28 V positions of the POWER connector, observing the 
polarities. 
8.5.4  External 9 to 28 VDC power supply connected to the internal terminals. 
1  –  Remove  all  connections  from  Connector  J15,  and  then  connect  the  additional  board,  Cat.  No., 
4005511  to  the  same  connector.  In  this case,  the  position  of  Switch  SW2  used  earlier  is  no  longer 
relevant. Switches SW1 and SW2 of the additional board will be used. 
2 - Create the following configuration:  
 
Switch SW1 of the additional board: Terminal 
 
Switch SW2 of the additional board: External 
3  –  Connect  the  external  power  supply  to  the  +/-  9  to  28  V  positions  of  the  POWER  connector, 
observing the polarities. 
 
 
 
 
 
 
 
 
 
 
Figure 31: Position to use for an external 9-28 V power supply