Valeport Ltd 0804IDU Manuel D’Utilisation

Page de 83
 
HydrINS 2 / HydrINS 2 Mini flow meter  
Installation Manual  
 
Page 51 / 83 
Route the cable using the cable gland, and then fasten the various wires to the  4-20 mA connector 
according to the correspondence table below: 
 
Terminal block (4-20 mA outputs) SDU only 
Function 
Screen 
Not Connected 
Common 
4-20mA GND 
Forward 
4-20 mA output normal direction 
Reverse 
4-20 mA output reverse direction 
Table 6: Connections for the active 4 – 20 mA outputs (power supplied by the display unit) 
 
If the cable has a braid, it can be connected to the Screen terminal of the connector in case of 
electrical interference. 
8.9.4  Wiring of the alarm outputs (Display A, C & G) 
The outputs are N-O (normally open) 
 
 
 
 
 
 
 
 
 
 
 
Terminal block Alarm outputs 
Function 
Screen 
Not Connected 
Alarm 1A 
Alarm 1 + 
Alarm 1B 
Alarm 1 - 
Alarm 2A 
Alarm 2 + 
Alarm 2B 
Alarm 2 -