Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S Manuel D’Utilisation
![Rigol](https://files.manualsbrain.com/attachments/4ef153c5b3a9e56ac40ea738ae281ea9dc808cdc/common/fit/150/50/a84f25c70c016e6b8d8e8d0fbe042c3bbe7b8a8feeee0ecbbc25ec79cb57/brand_logo.gif)
Codes de produits
MSO1074Z-S
RIGOL
DS1000Z Programming Guide
2-107
:TRIGger:RUNT:WLOWer
Syntax
:TRIGger:RUNT:WLOWer <NR3>
:TRIGger:RUNT:WLOWer?
:TRIGger:RUNT:WLOWer?
Description
Set or query the pulse width lower limit in runt trigger. The default unit is s.
Parameter
Name
Type
Range
Default
<NR3>
Discrete
8ns to 9.99s
1μs
Explanation
) is
GREater or GLESs.
Return
Format
The query returns the pulse width lower limit in scientific notation.
Example
:TRIGger:RUNT:WLOWer 0.01 /*Set the pulse width lower limit to 1ms*/
:TRIGger:RUNT:WLOWer? /*The query returns 1.000000e-03*/
Related
Command
:TRIGger:RUNT:ALEVel
Syntax
:TRIGger:RUNT:ALEVel <level>
:TRIGger:RUNT:ALEVel?
:TRIGger:RUNT:ALEVel?
Description
Set or query the trigger level upper limit in runt trigger. The unit is the same as the
current amplitude unit.
Parameter
Name
Type Range
Default
<level>
Real
(± 5 × VerticalScale from the screen center) -OFFSet 2V
Explanation
. For OFFSet, refer
Return
Format
The query returns the trigger level upper limit in scientific notation.
Example
:TRIGger:RUNT:ALEVel 0.16 /*Set the trigger level upper limit to 160mV*/
:TRIGger:RUNT:ALEVel? /*The query returns 1.600000e-01*/
Related
Command