Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S Manuel D’Utilisation
![Rigol](https://files.manualsbrain.com/attachments/4ef153c5b3a9e56ac40ea738ae281ea9dc808cdc/common/fit/150/50/a84f25c70c016e6b8d8e8d0fbe042c3bbe7b8a8feeee0ecbbc25ec79cb57/brand_logo.gif)
Codes de produits
MSO1074Z-S
RIGOL
DS1000Z Programming Guide
2-137
:WAVeform:YINCrement?
Syntax
:WAVeform:YINCrement?
Description
Query the waveform increment of the specified source in Y direction. The unit is the same
as the current amplitude unit.
Explanation
YINCrement= VerticalScale/25
Return
Format
The query returns the YINCrement in scientific notation.
Example
:WAVeform:YINCrement? /*The query returns 8.000000e-03*/
Related
Command
:WAVeform:YORigin?
Syntax
:WAVeform:YORigin?
Description
Query the vertical offset relative to the vertical reference position of the specified source
in Y direction.
Explanation
YORigin = VerticalOffset/YINCrement.
Return
Format
The query returns an integer.
Example
:WAVeform:YORigin? /*The query returns 0*/
Related
Command
:WAVeform:YREFerence?
Syntax
:WAVeform:YREFerence?
Description
Query the vertical reference position of the specified source in Y direction.
Explanation
The YREFerence is fixed at 127 (the bottom of the screen is 0 and the top is 255).
Return
Format
The query returns an integer.
Related
Command