Rigol DS1104Z 4-channel oscilloscope, Digital Storage oscilloscope, DS1104Z Manuel D’Utilisation
Codes de produits
DS1104Z
RIGOL
DS1000Z User’s Guide
1-10
Probe Compensation
When the probes are used for the first time, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorly
compensated probes may cause measurement inaccuracy or error. The probe
compensation procedures are as follows.
1. Perform steps 1, 2, 3 and 4 of “
” in the previous section.
2. Check the waveforms displayed and compare them with the following.
Figure 0-8 Probe Compensation
3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment
hole on the probe until the waveform displayed is as the “Perfectly
compensated” in the figure above.
compensated” in the figure above.
Over compensated Perfectly compensated Under compensated