Lecroy WAVEACE 2032 2-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2032 Manuel D’Utilisation
![Lecroy](https://files.manualsbrain.com/attachments/a6b425870750fec538437b1ab439f821a77f2154/common/fit/150/50/ff6974cbc973bf6214b742b12b227276bf1659fd6b79979bcdabfd971209/brand_logo.gif)
Codes de produits
WaveAce 2032
Operator's Manual
920836 Rev A
19
ACQUISITION - COUPLING, CPL
Description
The COUPLING command selects the coupling mode of the specified input channel.
The COUPLING? query returns the coupling mode of the specified channel.
Command Syntax
<channel>:CouPLing <coupling>
<channel>:= {C1,C2,C3,C4,M1,M2,M3,M4,M5,M6,M7,M8,M9,M10}
<coupling>:= {A1M*, D1M*, GND}
* Attenuation pertains only to instruments with a probe connected.
Query Syntax
<channel>:CouPLing?
Response Format
<channel>:CouPLing <coupling>
<coupling>:= {A1M, D1M, D50, GND, OVL}
COUPLING OVL is returned in the event of signal overload while in DC 50 Ω coupling. In this
condition, the oscilloscope will disconnect the input.
condition, the oscilloscope will disconnect the input.
Example
The following instruction sets the coupling of Channel 2 to Ground:
C2:CPL GND